14. TROUBLESHOOTING
Error Possible cause Actions
Press any key to turn indication on or
No display Energy-saving mode disable power-saving mode in the setup.
(Function setting mode)
(Power-save setting)
Wiring Carry out wire connection correctly according to
Supply voltage the connection point.
Setting of comparator Set a comparator output type (NPN, PNP)
output type to match the setup of apparatus.
No output (Function setting mode)
(Comparator operation selection)
Comparator enabled Enable a comparator in the setup.
(Comparator setting mode)
(Comparator Use/No use)
Display/output do Pressure leak Check pressure connection and piping.
not indicate when Verify applied pressure by using a pressure gauge
pressure is applied Too low pressure applied with the suitable pressure range.
(remains at zero)
Zero point shift Perform zero point adjustment
(zero point adjustment mode)
Pressure indication mode Select a pressure indication
(Function setting mode)
(Display setting)
Analog scaling mode Check analog output scaling setup
Unit display/output (Function setting mode)
(Analog Output Scaling)
conflicts with the Check the operational mode of the comparator, a
value of applied Set value of the preset value, and the preset value of delay time.
pressure comparator output (Function setting mode)
(Select comparator mode)
(Comparator setting mode)
(Comparator setting point)
(Comparator setting mode)
(Comparator delay)
Overloaded comparator Use unit with a load current set to 80mAdc or lower.
Foreign material Clean / remove obstruction in piping
(blockage)

E
rror
Display Contents Actions
Out of pressure display range (Upper limit)
FFF A
pressure above 110% FS of pressure
range is applied, or when indicated value
e
xceeds 1999. Adjust the applied pressure
O
ut of pressure display range (Lower limit)
within the rated pressure.
FFF A pressure less than –10% FS of pressure
r
ange is applied, or when indicated value
falls below –1999.
E-0 During zero point adjustment, applied
Open the unit to the atmos-
pressure is outside the range of ±10% FS
phere and perform zero
point adjustment again.