11
CASE
DISPLAY
CAP
Inside sensor Line
LCD holder
Notch
Transmission cable
(Twist)
Sheath
+
−
Terminal box
Display (board)
Shield
P
ower source
+ −
Receiver
T
ransmission cable
terminal strip, take care not to mishandle the display and associated
electronics.
• Turn display over to expose terminal strip, make positive and negative
• When connecting shield / drain wire, only connect one end which
should be at the receiver ground.
• Wiring stripping instructions, remove cable jacket 2-3˝ and strip wires
0.25˝. Shield / drain wire should not be exposed at the pressure
transmitter termination.
• Remove cover and carefully remove the display to access the