Note: Display not to be mounted in direct sunlight.
4. SPECIFICATION
Item: Description
A. Media / Wetted Parts:
Fluids or gases compatible with 316SS and 17-4PH SS
B. Overpressure (Proof):
–1500 psi FS and below; 200% FS value
–3000, 5000 psi; 150% FS value
–7500 psi; 120% FS value
C. Supply Voltage: 12-32 Vdc
D. Output:
4 to 20mA dc (two wire, output range: 3.2 to 20.8mA dc)
Response Time: 30 ms (with “0 filter setting)
Resolution: 0.1%F.S., Load resistance: 500ohms max.
E. Accuracy: ±0.25%F.S. (URL) includes the effects of linearity, hysteresis
and repeatability at 23°C (73°F) reference temperature
F. Temperature Effects
Operating and compensated temperature range (–10 to +60 °C);
±0.02%FS (URL)°C from 23°C reference.
G. Rangeability / Adjustment:
Zero: –10 to +110% FS
Span: –10 to + 110% FS
H. Display Accuracy: ±0.25%F.S. (URL) +1digit at 23°C
I. Display:
Character height: 10mm, LCD with LED backlight
Pressure / linear display: four LCD digits max.; 3
1
⁄2 digit
Display update: 500 ms
J. Units: Pressure Units: psi (2), (1) arbitrary
K. Setting Adjustments:
Internal key switches
Scaling function: Linear display / output
Max / Min function: Both the maximum and the mini-
mum readings can be indicated.
Filter function: User adjustable output damping, select from 0, 2,4, 8,
and 16 (0 = 30ms, 2 = 60ms, 4 = 120ms etc.)
Loop check function: User adjustable output for loop /
Vue de la page 6